Guide
How CBCT works: from X-ray cone to voxels
After this guide you will be able to explain in plain words how a cone beam CT scanner turns one rotation of an X-ray tube into a block of voxels, and why those numbers behave the way they do.
The source, the panel and the cone
A cone beam CT (CBCT) unit has two working parts mounted on a C shaped arm: an X-ray tube on one side and a flat panel detector on the other. The patient sits or stands between them, head held still. During the scan the arm swings around the head, through a half turn or a full turn, and the tube fires short pulses as it goes.
The beam is shaped like a cone, wide enough to cover the whole region of interest at once. A medical CT instead sweeps a thin fan of X-rays a slice at a time (see CBCT vs CT scan). The cone covers a cylinder of anatomy in one pass, and that cylinder is the field of view (see field of view and voxel size).
What one projection records
Each pulse produces one projection: a two dimensional image on the flat panel, like a radiograph from that angle. The panel is a grid of small detector elements, each measuring how much of the beam reached it. Little arrives where the ray crossed enamel or thick cortical bone, so that pixel is bright on the final image; almost everything arrives where it crossed air.
A projection therefore records the total attenuation along each ray. A molar root, the buccal plate, the tongue and the far side of the jaw all land on the same pixel if they share a ray. One projection cannot tell you where along that ray the attenuation happened. That is the limit of every plain radiograph (see CBCT vs periapical radiograph), and it is the problem that reconstruction solves.
From projections to voxels: back projection
During one rotation the scanner collects a few hundred projections from evenly spaced angles. Reconstruction runs them backwards. Divide the field of view into a grid of tiny cubes, the voxels, all set to zero. Smear the first projection's values back along the rays that produced it, adding a share to every voxel each ray crossed. Repeat for every other projection from its own angle.
A voxel that really contains dense material is crossed by a high reading in every projection, so it accumulates a large value. A voxel of air is crossed by low readings from every direction and stays dark. Smearing on its own produces a blurred result, so each projection is first passed through a sharpening filter, which is why the method is called filtered back projection. The cone beam version, published by Feldkamp, Davis and Kress, also corrects for the tilt of the rays at the top and bottom of the cone.
The output is a three dimensional array of numbers. In the demo scan it is 480 x 480 x 320 voxels, about 74 million of them, stored as a DICOM series (see what is DICOM). Every slice you scroll in the viewer is one layer of that array; every reformat is the same array read in a different direction (see tri-planar MPR).
Why voxels are isotropic
A medical CT sets its slice thickness and its pixel size separately, and they are often different. A cone beam samples all three directions in one rotation with the same square detector elements, so no direction has reason to be coarser than another. Reconstruction naturally produces cubes: voxels of the same size in every direction, which is what isotropic means.
The voxel size depends on the field of view and the resolution setting, typically a few tenths of a millimetre and smaller for a small field. The demo scan's voxels are 0.266 x 0.266 x 0.252 mm, close to a cube. Isotropic voxels are why a coronal or sagittal reformat looks as sharp as the axial slice, and why you can measure in any plane with the same confidence (see axial, coronal and sagittal planes).
Why grey values are not calibrated
A medical CT converts reconstructed attenuation into Hounsfield units, a scale on which water reads 0 and air about minus 1000 on any machine. A CBCT cannot do this reliably, for three reasons that all come from the cone.
First, a wide beam produces a lot of scattered radiation, and scatter reaching the panel is counted as if it had come straight through, which flattens the measured values. Second, the field of view is smaller than the head, so anatomy outside it attenuates the beam without being reconstructed, and its effect is spread over the voxels inside. Third, the beam is a mix of energies and the softer ones are absorbed first, so the same tissue reads differently depending on what the ray has already crossed. The result varies with position in the field, with the neighbouring anatomy and with the machine. JawAtlas therefore shows intensities as grey values and never as HU (see window, level and grey values).
Where artifacts come from
Most CBCT artifacts are the reconstruction doing exactly what it was told with imperfect projections. If the patient moves between projections, the rays no longer agree about where things are, and edges double or blur. If a metal crown absorbs almost the whole beam, the projections through it carry no information, and the back projection smears that missing data into dark bands and bright streaks radiating from the metal. Beam hardening darkens the bone between two dense objects. A faulty detector element appears in every projection at the same place and reconstructs as a ring. At the top and bottom of the field the tilted rays of the cone are sampled less completely, so those slices are blurrier and noisier than the middle. The artifacts guide shows each on real slices.
Try it on the demo scan
Open the demo scan and go to the 3D stage. Switch on MIP and drag to rotate: these are projections through the reconstructed voxels, the nearest thing in the app to what the panel captured, and the roots of both sides overlap and separate as the angle changes (see maximum intensity projection). Then go to the MPR stage and scroll the axial stack with the vertical slider to the top slices: compared with the middle of the volume they are noisier, which is the cone beam effect. Finally use Adjust mode to drag the window and watch the readout, for example "C 300 / W 1500 grey", change: those are grey values of this one scan, not calibrated units.
Questions students ask
How does a CBCT scanner actually work?
An X-ray tube and a flat panel detector swing once around the patient's head. At each of a few hundred angles the tube fires a pulse and the panel records a two dimensional projection, like a radiograph from that angle. A computer then back projects all of them into a three dimensional grid of voxels, which is the volume you scroll through in a viewer.
What is a voxel in CBCT?
A voxel is one cell of the reconstructed volume, a tiny cube with a single grey value. In CBCT the voxels are usually the same size in all three directions, typically a few tenths of a millimetre, which is why coronal and sagittal reformats look as sharp as the axial slices.
Why are CBCT grey values not Hounsfield units?
The wide cone beam collects a great deal of scatter, anatomy outside the field of view attenuates the beam without being reconstructed, and beam hardening shifts values near dense objects. The result is a number that changes with position, neighbouring anatomy and machine, so it cannot be calibrated the way a medical CT's Hounsfield scale is.
Why does CBCT have artifacts?
Reconstruction assumes every projection agrees about where things are and how dense they are. Patient movement, metal that blocks the beam, a mix of X-ray energies and imperfect detector elements all break that assumption, and the back projection turns the disagreement into blur, streaks, dark bands or rings.